ResiQuant® Ready-to-use E.coli DNA Residue Detection Kit
|
Performance
|
Result
|
1. Linear
|
Quantitative range
|
30 fg/μL-300 pg/μL
|
PASS
|
R2
|
>0.98
|
PASS
|
Amplification efficiency
|
90%~110%
|
PASS
|
2. Accuracy
|
Deviation from standard
|
CV≤15%
|
PASS
|
3. Recovery
|
Simulated sample recovery rate
|
70%~130%
|
PASS
|
4. Precision
|
Repeatability
|
CV≤15%
|
PASS
|
Intermediate precision
|
CV≤30%
|
PASS
|
5. Exclusivity
|
No interference from other DNA
|
PASS
|
6. Limit of quantitation
|
30 fg/μl
|
PASS
|
7. Bank limit
|
The bank sample was retested 24 times, and no less than 23 reactions had Ct≥36 or went not detected.
|
PASS
|
8. Durability
|
DNA fragmentation
|
Test results are not affected
|
PASS
|
Instrument suitability
|
ABI/Agilent/Bio-Rad
|
PASS
|
9. Stability
|
Repeated freezing and thawing
|
Detection performance is not affected by repeated freezing and thawing 5 times
|
PASS
|